Search results for "Flexisource Co-60"

showing 1 items of 1 documents

Monte Carlo dosimetric study of the Flexisource Co-60 high dose rate source.

2011

Purpose: Recently, a new HDR 60 Co brachytherapy source, Flexisource Co-60, has been developed (Nucletron B.V. Veenendaal, The Netherlands). This study aims to obtain dosimetric data for this source for its use in clinical practice as required by AAPM and ESTRO. Material and methods: Two Monte Carlo radiation transport codes were used: Penelope2008 and GEANT4. The source was centrally-positioned in a 100 cm radius water phantom. Absorbed dose and collisional kerma were obtained using 0.01 cm (close) and 0.1 cm (far) sized voxels to provide high-resolution dosimetry near (far from) the source. Dose rate distributions obtained with the two Monte Carlo codes were compared. Results and Discussi…

business.industryFlexisource Co-60Penelope2008medicine.medical_treatmentBrachytherapyMonte Carlo methodbrachytherapyImaging phantomComputational physicsKermaOncologyAbsorbed doseMedicineDosimetryRadiology Nuclear Medicine and imagingOriginal ArticlebusinessDose rateRadiation treatment planningNuclear medicineTG-43GEANT4Journal of contemporary brachytherapy
researchProduct